Temperature reference probes

 

Temperature reference probe:

Technical specifications:

- Resistance : Nominal 100 ohm ± 0.1 ohm at 0°C
- Temperature range: -200 to 500 °C
- Sheath materiel : inconel 600
- Hysteresis < 0,01 °C at 0°C using -183 and 420 °C as the end points
- Probe accuracy

- ± 0,018 °C @ -196°C
- ± 0,018 °C @ 0°C
- ± 0,019°C @ 200°C
- ± 0,023°C @ 420°C

- Size: type 5612 : 0,187 " dia. x 9 " ( 0,475 x 22,86 cm)
type 5613 : 0,187 " dia. x 6 " ( 0,475 x 15,24 cm)
type 5614 : 0,25 " dia x 12 " ( 0,635 x 30,48 cm)
- manufacturer : HART

Smart thermometer :

- Temperature range : -200 to 960 °C
- Resistance range : 0 to 400 ohm
- Probe : 25 ou 100 ohm
- Characterisation : ITTS-90, IPTS-68 or Callendar-Van Dusen
- Temperature accuracy :
- ± 0,006°C @ 0°C
- ± 0,009°C @ 100°C
- ± 0,012°C @ 200°C
- ± 0,018°C @ 400°C
- ± 0,024°C @ 600°C

- Temperature coefficient : <0,5 ppm/°C, typical
- Temperature resolution : 0,001°C
- Resistance resolution : 0,0001 ohm
- Calibration : Coefficients are entered directly in the equipment during the annual calibrations. ITS-90 formulas reside in its firmware
The display will indicate the true temperature value
- Display : 8 digit, 7 segment LED
- Size : 14,3 x 18,1 x 6,1 cm
- Weight : 1 Kg
- Manufacturer : HART

Software XpertRTD:

The complete system is delivered with XpertRTD software of LIVES International for:
- displaying and storing simultaneously up to 9 temperature reference probes ( HART or Smart Probes)
- generate Excel reports with data provided by the 9 probes
- saving time and money for sterilisation equipment probe's calibration
- manufacturer: Lives International