GMP SW

By Lives International Tuesday, 20 September 2016 00:00
Rate this item
(1 Vote)

Xpertlog® GMP simplifies and makes the thermal validation process more efficient:
High efficiency: easy to program and install, pre-calibration not needed
High reliability: no data loss during validation process
High accuracy: ±0.1°C from -90°C to +140°C.
Data integrity: FDA 21 CFR Part 11

Characteristics GMP SW  
Sensor element 4 wires PT-100
- rigid - 20mm (0.8")
Operating Range - 40 °C to 140 °C
Accuracy ±0.1 °C from -40 °C to 140 °C
Memory capacity 32,000 data points temperature
Sampling rate 1 sec to 24 hours
Internal clock drift 4 sec / 24 h @ 23 °
Mechanical
- material Peek 1000 - FDA compliant
- enclosure IP-68
- weight 91g (3.21 oz)
- diameter 50 mm (1.97")
- height 40 mm (1.57")
Power 1 Lithium battery
0.35 g Lithium content
classified UN3091 under IATA DGR section 3.9.2.6
compliant with ICAO/IATA packing instructions 970
user replaceable
Battery lifetime Up to 6 months, based on usage
Docking station USB connection
4, 8, 12 and 20 ports available
programs and reads unlimited numbers of loggers
Calibrations Factory:
User:
traceable NIST/COFRAC - ITS-90 coefficients stored in the internal memory
close loop calibration, using XpertLog® software
Certification FMC:
Health Safety:
EN 60601-1-2 / 2007
EN 60950-1 / 2006
  • GMP
    GMP Standard data loggers for recording temperature values. Easy to program and install, no data loss during validation process, multiple users…
  • GMPL
    GMPL Standard data logger for recording temperature values with long rigid sensor. Easy to program and install, no data loss during…
  • GMPF
    GMPF Standard data logger with long flexible sensor for recording temperature values. Easy to program and install, no data loss during…
  • GMPF LP
    GMPF LP Low profile (35mm height) standard data logger with long flexible sensor for recording temperature values. Easy to program and install,…
  • GMP SW
    GMP SW Standard data logger for capturing humidity and temperature values. Easy to program and install, highly reliable (no data loss), high…
  • GMP RH
    GMP RH Standard data loggers for recording humidity and temperature values. Easy to program and install, no data loss during validation process,…
  • GMPF RH
    GMPF RH Standard data loggers with flexible temperature sensor and rigid humidity sensor. Easy to program and install, no data loss during…
  • GMP P
    GMP P Standard data loggers for recording pressure and temperature values. Easy to program and install, no data loss during validation process,…
  • GMPF P
    GMPF P Standard data logger for recording pressure and temperature values, with flexible temperature sensor. Easy to program and install, no data…
  • Pliers
    Pliers Tools to open loggers for battery replacemenet.
  • Bracket
    Bracket Mounting bracket for standard and RF data loggers
  • Docking Station
    Docking Station Docking Station for GMP and GDP data loggers
ISO 9001:2008

ISO 9001:2008

ISO 14001:2004

ISO 14001:2004

OHSAS 18001:2007

ISO 18001:2007

Read 320 times
More in this category: « GMPF LP GMP RH »

Leave a comment

Offices

USA
1 Chestnut Street
Nashua, NH 03063
CANADA
7240 Waverly, Suite 209
Montreal QC H2R 2Y8
CHINA
1511, Building B, No.15 Changyi Road,
Shanghai 200439
UNITED KINGDOM
235-237 Vauxhall Bridge Road
SW1V 1EJ London
FRANCE
3 rue Jules Guesde
92300 Levallois Perret
ROMANIA
37 Bela Bartok
Dumbravita 307160, jud. Timis
GERMANY
15 Carl-Legien-Str.
Gebäude B, 63073 Offenbach-Waldhof

UPCOMING EVENTS