Page 13 - Lives International Product Catalog 2019
P. 13
GMPF RF 30H
Real Time Temperature Data Logger -90 to +140°C
+ Collects and transmits real time temperature to XpertLog Software
+ ITS-90 calibration coefficients are stored in the internal memory
+ Complies with FDA 21 CFR Part 11
+ Network ready: multiple users, centralized, server database, active directory authentication
+ Calibration can be performed on site
+ Can be remotely activated
ø 37.00mm
1.46″
62.81mm 2.47″ ø 20.00mm
0.79″
Sensor type 4 wires / 4 wires thin
- flexible 4 wires: 480 mm (19”) / 4 wires thin: 150 mm (6") or custom length
- sensor size 4 wires: 1.5 x 8 mm (0.06 x 0.315”)
4 wires thin: 1.2 x 2.8 mm (0.05 x 0.11”)
- cross section 4 wires: 1.77 mm (0.0027 inch )
2
2
2
2
4 wires thin: 1.13 mm (0.00175 inch )
Operating Range Order Reference:
- body temp. -90°C to +140°C
- sensor temp. 4 wires: -200°C to +200°C GMPF_RF_30H_4W
4 wires thin: -90°C to +140°C
Accuracy
- 4 wires ±0.1°C from -90 to +140°C GMPF_RF_30H_4W_THIN
- 4 wires thin ±0.1°C from -90 to +140°C
Resolution 0.01°C
Memory capacity 32,000 data points
Sampling rate 1 sec to 24 hours
Transmission Simultaneous: maximum 20 loggers
Sequential: unlimited
Internal clock Drift 4 sec / 24 h @ + 23 °C Standard Calibration Points
Mechanical
- material Peek 1000 - FDA compliant • -80°C
- enclosure IP-68
- weight 50 g (1.76 oz)
- diameter 37 mm (1.46”) • 0°C
- height 62.81 mm (2.47”)
Power 2 x Lithium batteries, 0.35 g/each Lithium content classified • 140°C
UN3091 under IATA DGR section 3.9.2.6 compliant with
ICAO/IATA packing instruction 970 user replaceable
Battery lifetime Up to 6 months, based on usage
Antenna Internal
Wireless range 121 m (400’)
Radio frequency 2.4 GHz ISM band
Access point USB to computer
Programs and reads unlimited number of loggers
Calibrations Factory: Traceable NIST / COFRAC - ITS-90 coefficients reside in the internal memory
®
User: Closed loop user calibration, using XpertLog Software
Certification EMC: EN 60601-1-2 / 2007
RF: EN 300440-2 v1.3.1 (2009-03)
Health & Safety: EN 60950-1/2006 GMPF RF 30H
-13-