GMPF P RF
Real Time Pressure Data Logger 0.01 to 5.0 BarA
+ Collects and transmits real time pressure and temperature to XpertLog Software
+ ITS-90 calibration coefficients are stored in the internal memory
+ Complies with FDA 21 CFR Part 11
+ Network ready: multiple users, centralized, server database, active directory authentication
+ Can be remotely activated
Sensor type Absolute pressure / Temp: 4 wires PT-100
- flexible 480 mm (19”) PTFE/SS
- customized Up to 5000 mm (195”)
- temp. sensor size 1.5 x 8 mm (0.06 x 0.315”) Order Reference:
- temp. cross section 1.77 mm (0.0027 inch )
2
2
Operating Range GMPF-P_RF
- pressure 0.01 to 5 BarA
- body temp. -40°C to +140°C
- sensor temp. -200°C to +200°C
Accuracy ±10 mbar ≤ 3 BarA @ 130°C; ±20 mbar ≤ 3 BarA @ 23°C;
±30 mbar > 3 BarA @ 130°C; ±50 mbar > 3 BarA @ 23°C;
±0.1°C from -40 to +140°C
Resolution 0.1mbar; 0.01°C Standard Calibration Points
Memory capacity 16,000 data points pressure / 16,000 data points temperature
Sampling rate 1 sec to 24 hours hysteresis at 23°C and 130°C:
Transmission Simultaneous: maximum 20 loggers • 0.05 BarA
Sequential: unlimited • 0.50 BarA
Internal clock Drift 4 sec / 24 h @ + 23 °C
Mechanical • 1.00 BarA
- material Peek 1000 - FDA compliant • 1.50 BarA
- enclosure IP-68 • 2.00 BarA
- weight 129 g (4.55 oz) • 2.50 BarA
- diameter 64.50 mm (2.54”)
- height 51.0 mm (2.01”) • 3.00 BarA
Power 2 x Lithium batteries, 0.35 g/each content classified • 3.50 BarA
UN3091 under IATA DGR section 3.9.2.6 compliant with • 4.00 BarA
ICAO/IATA packing instruction 970 user replaceable • 4.50 BarA
Battery lifetime Up to 6 months, based on usage
Antenna Internal • 5.00 BarA
Wireless range 121 m (400’)
Radio frequency 2.4 GHz ISM band
Access point USB to computer
Programs and reads unlimited number of loggers
Calibrations Factory: Traceable NIST / COFRAC - ITS-90 coefficients reside in the internal memory
®
User: Closed loop user calibration, using XpertLog Software
GMPF P RF
Certification EMC: EN 60601-1-2 / 2007
RF: EN 300440-2 v1.3.1 (2009-03)
Health & Safety: EN 60950-1/2006
-19-