P. 19

GMPF P RF






                          Real Time Pressure Data Logger  0.01 to 5.0 BarA







                                 +    Collects and transmits real time pressure and temperature to XpertLog Software
                                 +    ITS-90 calibration coefficients are stored in the internal memory
                                 +    Complies with FDA 21 CFR Part 11
                                 +    Network ready: multiple users, centralized, server database, active directory authentication
                                 +    Can be remotely activated
















              Sensor type        Absolute pressure / Temp: 4 wires  PT-100
                 - flexible      480 mm (19”) PTFE/SS
                 - customized    Up to 5000 mm (195”)
                 - temp. sensor size   1.5 x 8 mm (0.06 x 0.315”)                              Order Reference:
                 - temp. cross section   1.77 mm  (0.0027 inch )
                                                 2
                                       2
              Operating Range                                                                  GMPF-P_RF
                 - pressure      0.01 to 5 BarA
                 - body temp.    -40°C to +140°C
                 - sensor temp.    -200°C to +200°C
              Accuracy           ±10 mbar ≤ 3 BarA @ 130°C;   ±20 mbar ≤ 3 BarA @ 23°C;
                                 ±30 mbar > 3 BarA @ 130°C;   ±50 mbar > 3 BarA @ 23°C;
                                 ±0.1°C from -40 to +140°C
              Resolution         0.1mbar; 0.01°C                                      Standard Calibration Points
              Memory capacity    16,000 data points pressure / 16,000 data points temperature
              Sampling rate      1 sec to 24 hours                                       hysteresis at 23°C and 130°C:
              Transmission       Simultaneous: maximum 20 loggers                           • 0.05 BarA
                                 Sequential: unlimited                                      • 0.50 BarA
              Internal clock Drift    4 sec / 24 h @ + 23 °C
              Mechanical                                                                    • 1.00 BarA
                 - material      Peek 1000 - FDA compliant                                  • 1.50 BarA
                 - enclosure     IP-68                                                      • 2.00 BarA
                 - weight        129 g (4.55 oz)                                            • 2.50 BarA
                 - diameter      64.50 mm (2.54”)
                 - height        51.0 mm (2.01”)                                            • 3.00 BarA
              Power              2 x Lithium batteries, 0.35 g/each content classified      • 3.50 BarA
                                 UN3091 under IATA DGR section 3.9.2.6 compliant with       • 4.00 BarA
                                 ICAO/IATA packing instruction 970 user replaceable         • 4.50 BarA
              Battery lifetime   Up to 6 months, based on usage
              Antenna            Internal                                                   • 5.00 BarA
              Wireless range     121 m (400’)
              Radio frequency    2.4 GHz ISM band
              Access point       USB to computer
                                 Programs and reads unlimited number of loggers
              Calibrations       Factory:              Traceable NIST / COFRAC - ITS-90 coefficients reside in the internal memory
                                                                              ®
                                 User:                    Closed loop user calibration, using XpertLog  Software
                                                                                                                       GMPF P RF
              Certification      EMC:                     EN 60601-1-2 / 2007
                                 RF:                       EN 300440-2 v1.3.1 (2009-03)
                                 Health & Safety:      EN 60950-1/2006







                                                                                                                   -19-


                   LVSC.3.1.6 - October 2019                                        Visit www.lives-international.com for more details.