Page 24 - Lives International Product Catalog 2019
P. 24

GMPF






                           Standard Temperature Data Logger  -90 to +140°C











                               +    Collects and stores temperature information in the internal memory
                               +    Recover the data through the docking station using XpertLog Software
                               +    ITS-90 calibration coefficients are stored in the internal memory
                               +    Complies with FDA 21 CFR Part 11
                               +    Network ready: multiple users, centralized, server database, active directory authentication
                               +    Calibration can be performed on site






                             23.00 mm    0.91″                      16.00mm
                                                                      0.63″
                                                                      40.00mm
                                                                        1.57″
                             27.00 mm     1.06″  50.00mm / 1.97″





               Sensor type        4 wires  PT-100
                  - flexible      480 mm (19”) PTFE/SS
                  - customized    Up to 5000 mm (195”)
                  - sensor size   1.5 x 8 mm (0.06 x 0.315”)
                  - cross section   1.77 mm  (0.0027 inch )                               Order Reference:
                                                  2
                                        2
               Operating Range
                  - body temp.    -90°C to +140°C
                  - sensor temp.   -200°C to +200°C                                       GMPF
               Accuracy           ±0.1°C from -90 to +140°C
               Resolution         0.01°C
               Memory capacity    32,000 data points
               Sampling rate      1 sec to 24 hours                             Standard Calibration Points
               Internal clock Drift    4 sec / 24 h @ + 23 °C
               Mechanical                                                              • -80°C
                  - material      Peek 1000 - FDA compliant
                  - enclosure     IP-68
                  - weight        44 g (1.55 oz)                                       • 0°C
                  - diameter      40.0 mm (1.57”)
                  - height        50.0 mm (1.97”)                                      • 140°C
               Power              1 x Lithium battery, 0.35 g Lithium content classified
                                  UN3091 under IATA DGR section 3.9.2.6 compliant with
                                  ICAO/IATA packing instruction 970 user replaceable
               Battery lifetime   Up to 12 months, based on usage
               Docking station    USB to computer
                                  4, 8, 12 and 20 ports available
                                  Programs and reads unlimited number of loggers
               Calibrations       Factory:               Traceable NIST / COFRAC - ITS-90 coefficients reside in the internal memory
                                  User:                    Closed loop user calibration, using XpertLog  Software
                                                                               ®
               Certification      EMC:                    EN 60601-1-2 / 2007
      GMPF
                                  Health & Safety:      EN 60950-1/2006










   -24-
   19   20   21   22   23   24   25   26   27   28   29