Page 28 - Lives International Product Catalog 2019
P. 28

GMP P








                              Standard Pressure Data Logger 0.01-5.0 BarA






                              +    Collects and stores pressure and temperature information in the internal memory
                              +    Recover the data through the docking station using XpertLog Software
                              +    ITS-90 calibration coefficients are stored in the internal memory
                              +    Complies with FDA 21 CFR Part 11
                              +    Network ready: multiple users, centralized, server database, active directory authentication








                           69.00 mm    2.72″                          30.00mm / 1.18″
                               96.00mm / 3.78″                        40.00mm / 1.57″





                           27.00 mm     1.06″



                Sensor type        Absolute pressure / Temp: 4 wires  PT-100
                   - rigid         20 mm (0.8”)
                   - temp. sensor size   1.5 x 8 mm (0.06 x 0.315”)
                   - temp. cross section   1.77 mm  (0.0027 inch )
                                         2
                                                  2
                Operating Range                                                             Order Reference:
                   - pressure      0.01 BarA to 5.0 BarA
                   - body temp.    -40°C to +140°C
                   - sensor temp.   -40°C to +140°C                                         GMP-P
                Accuracy           ±10 mbar ≤ 3 BarA @ 130°C;    ±30 mbar > 3 BarA @ 130°C
                                   ±20 mbar ≤ 3 BarA @ 23°C;     ±50 mbar > 3 BarA @ 23°C
                                   ±0.1°C from -40 °C to 140 °C
                Resolution         0.1 mbar; 0.01°C                                Standard Calibration Points
                Memory capacity    16,000 data points pressure / 16,000 data points temperature         hysteresis at 23°C and 130°C:
                Sampling rate      1 sec to 24 hours
                Internal clock Drift    4 sec / 24 h @ + 23 °C                            • 0.05 BarA
                Mechanical                                                                • 0.50 BarA
                   - material      Peek 1000 - FDA compliant                              • 1.00 BarA
                   - enclosure     IP-68                                                  • 1.50 BarA
                   - weight        77 g (2.72 oz)
                   - diameter      40.0 mm (1.57”)                                        • 2.00 BarA
                   - height        96.0 mm (3.78”)                                        • 2.50 BarA
                Power              1 x Lithium battery, 0.35 g Lithium content classified     • 3.00 BarA
                                   UN3091 under IATA DGR section 3.9.2.6 compliant with     • 3.50 BarA
                                   ICAO/IATA packing instruction 970 user replaceable
                Battery lifetime   Up to 12 months, based on usage                        • 4.00 BarA
                Docking station    USB to computer                                        • 4.50 BarA
                                   4, 8, 12 and 20 ports available                        • 5.00 BarA
                                   Programs and reads unlimited number of loggers
                                   Factory:               Traceable NIST / COFRAC - ITS-90 coefficients reside in the internal memory
                Calibrations
     GMP P      Certification      User:                    Closed loop user calibration, using XpertLog  Software
                                                                               ®
                                   EMC:                    EN 60601-1-2 / 2007
                                   Health & Safety:      EN 60950-1/2006







  -28-