Page 29 - Lives International Product Catalog 2019
P. 29
GMPF P
Standard Pressure Data Logger 0.01 to 5.0 BarA
+ Collects and stores pressure and temperature information in the internal memory
+ Recover the data through the docking station using XpertLog Software
+ ITS-90 coefficients are stored in the internal memory
+ Complies with FDA 21 CFR Part 11
+ Network ready: multiple users, centralized, server database, active directory authentication
68.00 mm / 2.68″ 96mm / 3.78″ 30.00mm / 1.18″
40.00mm / 1.57″
28.00 mm 1.10″
Sensor type Absolute pressure / Temp: 4 wires PT-100
- flexible 480 mm (19”)
- customized Up to 5000 mm (195”)
- temp. sensor size 1.5 x 8 mm (0.06 x 0.315”)
- temp. cross section 1.77 mm (0.0027 inch ) Order Reference:
2
2
Operating Range
- pressure 0.01 to 5.0 BarA GMPF-P
- body temp. -40°C to +140°C
- sensor temp. -200°C to +200°C
Accuracy ±10 mbar ≤ 3 BarA @ 130°C; ±20 mbar ≤ 3 BarA @ 23°C
±30 mbar > 3 BarA @ 130°C; ±50 mbar > 3 BarA @ 23°C Standard Calibration Points
±0.1°C from -40 °C to 140 °C
Resolution 0.1 mbar; 0.01°C hysteresis at 23°C and 130°C:
Memory capacity 16,000 data points pressure / 16,000 data points temperature • 0.05 BarA
Sampling rate 1 sec to 24 hours
Internal clock Drift 4 sec / 24 h @ + 23 °C • 0.50 BarA
Mechanical • 1.00 BarA
- material Peek 1000 - FDA compliant • 1.50 BarA
- enclosure IP-68 • 2.00 BarA
- weight 77 g (2.72 oz)
- diameter 40.0 mm (1.57”) • 2.50 BarA
- height 96.0 mm (3.78”) • 3.00 BarA
Power 1 x Lithium battery, 0.35 g Lithium content classified • 3.50 BarA
UN3091 under IATA DGR section 3.9.2.6 compliant with • 4.00 BarA
ICAO/IATA packing instruction 970 user replaceable
Battery lifetime Up to 12 months, based on usage • 4.50 BarA
Docking station USB to computer • 5.00 BarA
4, 8, 12 and 20 ports available
Programs and reads unlimited number of loggers
Calibrations Factory: Traceable NIST / COFRAC - ITS-90 coefficients reside in the internal memory
®
User: Closed loop user calibration, using XpertLog Software
Certification EMC: EN 60601-1-2 / 2007
Health & Safety: EN 60950-1/2006 GMPF P
-29-
LVSC.3.1.6 - October 2019 Visit www.lives-international.com for more details.