XpertVal






                                      Thermocouple systems






                                        +    Conventional Tc’s based system for thermal mapping and validation
                                        +    60 (up to 120) channels simultaneously
                                        +    Rugged data logger used in industrial harsh environment, real time data display
                                        +    High accuracy  Tc after pre-calibration (±0.2°C)
                                        +    Complies with FDA 21 CFR Part 11
                                        +    Network ready: multiple users, centralized, server database










                Sensors            Thermocouples
                Sensor length       - standard:    10 m (33 feet)
                                    - customizable:      custom length
                Sensor type        All type of thermocouples, most used : T, J, K
                Temperature range   - teflon:       -200°C to +200°C
                                   - kapton:      -200°C to +400°C                              Order Reference:
                                   - others:      up to +1200°C
                Accuracy          ±0.2°C after pre-calibration                                  Lives XpertVal 60
                Memory capacity    Internal memory:  20 Mb
                                   Backup memory: USB storage device compatible
                Tc Tip             Teflon: teflon tip to prevent moisture on the wire
                                   Kapton: no tip
                Sampling rate      120 channels - 5 sec
                                     60 channels - 5 sec
                                     20 channels - 2 sec
                                     12 channels - 1 sec
                Software           21 CFR Part 11
                Module             Universal input module:    4 - 20 mA / 0-10 V / RTD (PT-100  2, 3 or 4 wires), thermocouples
                Communication      USB XpertVal to PC
                                   LAN
                                   2.4 GHz, 802.11b/g/n
                Close loop calibration   Required when using thermocouples for thermal validation
                                   System uses temperature block and reference probe, software controlled
                                   reports generated automatically by the software
                Reference probes   RTD, Fluke 1502, Lives XpertRTD
                Dry blocks         Lives :   PTC 125 : -90°C to +125°C    /    RTC 157 : -45°C to +155°C    /    RTC 159 : -100°C to +155°C
                                                  RTC 700 : +23°C to +700°C
                                   Fluke 9190: -90°C to +140°C; Fluke 9170: -45°C to +140°C; Fluke 9172: +35°C to +425°C
                Calibrations       Factory:   NIST / COFRAC
                                                             ®
                Users              Close loop calibration using XpertLog  software
                 XpertVal® Data Logger      Reference Probe            Ametek Dry Blocks          Fluke Dry Blocks
                                              Fluke 1502







     XperVal       Thermocouples       Pressure Transducer  Wireless Module     Reference Probe    Fluke Oil Baths


                                                                                   Xpert RTD







  -36-


                        Visit www.lives-international.com for more details.                                        LVSC.3.1.6 - October 2019
   31   32   33   34   35   36   37   38   39   40   41