Page 36 - Lives International Product Catalog 2019
P. 36
XpertVal
Thermocouple systems
+ Conventional Tc’s based system for thermal mapping and validation
+ 60 (up to 120) channels simultaneously
+ Rugged data logger used in industrial harsh environment, real time data display
+ High accuracy Tc after pre-calibration (±0.2°C)
+ Complies with FDA 21 CFR Part 11
+ Network ready: multiple users, centralized, server database
Sensors Thermocouples
Sensor length - standard: 10 m (33 feet)
- customizable: custom length
Sensor type All type of thermocouples, most used : T, J, K
Temperature range - teflon: -200°C to +200°C
- kapton: -200°C to +400°C Order Reference:
- others: up to +1200°C
Accuracy ±0.2°C after pre-calibration Lives XpertVal 60
Memory capacity Internal memory: 20 Mb
Backup memory: USB storage device compatible
Tc Tip Teflon: teflon tip to prevent moisture on the wire
Kapton: no tip
Sampling rate 120 channels - 5 sec
60 channels - 5 sec
20 channels - 2 sec
12 channels - 1 sec
Software 21 CFR Part 11
Module Universal input module: 4 - 20 mA / 0-10 V / RTD (PT-100 2, 3 or 4 wires), thermocouples
Communication USB XpertVal to PC
LAN
2.4 GHz, 802.11b/g/n
Close loop calibration Required when using thermocouples for thermal validation
System uses temperature block and reference probe, software controlled
reports generated automatically by the software
Reference probes RTD, Fluke 1502, Lives XpertRTD
Dry blocks Lives : PTC 125 : -90°C to +125°C / RTC 157 : -45°C to +155°C / RTC 159 : -100°C to +155°C
RTC 700 : +23°C to +700°C
Fluke 9190: -90°C to +140°C; Fluke 9170: -45°C to +140°C; Fluke 9172: +35°C to +425°C
Calibrations Factory: NIST / COFRAC
®
Users Close loop calibration using XpertLog software
XpertVal® Data Logger Reference Probe Ametek Dry Blocks Fluke Dry Blocks
Fluke 1502
XperVal Thermocouples Pressure Transducer Wireless Module Reference Probe Fluke Oil Baths
Xpert RTD
-36-
Visit www.lives-international.com for more details. LVSC.3.1.6 - October 2019