Page 4 - Lives International Product Catalog 2019
P. 4
GMPF RF
Real Time Temperature Data Logger -90 to +140°C
+ Collects and transmits real time temperature to XpertLog Software
+ ITS-90 calibration coefficients are stored in the internal memory
+ Complies with FDA 21 CFR Part 11
+ Network ready: multiple users, centralized, server database, active directory authentication
+ Calibration can be performed on site
+ Can be remotely activated
21.50mm 0.85″ 65.00mm
2.56″
15.50mm
33.50mm 1.32″ 55.00mm 2.17″ 0.61″
Sensor type 4 wires PT-100
- flexible 480 mm (19”)
- customized Up to 5000 mm (195”)
- sensor size 1.5 x 8 mm (0.06 x 0.315”)
- cross section 1.77 mm (0.0027 inch )
2
2
Operating Range
- body temp. -90°C to +140°C
- sensor temp. -200°C to +200°C Order Reference:
Accuracy ±0.1°C from -90 to +140°C
Resolution 0.01°C GMPF_RF
Memory capacity 32,000 data points
Sampling rate 1 sec to 24 hours
Transmission Simultaneous: maximum 20 loggers
Sequential: unlimited
Internal clock Drift 4 sec / 24 h @ + 23 °C
Mechanical
- material Peek 1000 - FDA compliant Standard Calibration Points
- enclosure IP-68
- weight 115 g (4.06 oz)
- diameter 65 mm (2.56”) • -80°C
- height 55 mm (2.17”)
Power 2 x Lithium batteries, 0.35 g/each content classified • 0°C
UN3091 under IATA DGR section 3.9.2.6 compliant with
ICAO/IATA packing instruction 970 user replaceable • 140°C
Battery lifetime Up to 6 months, based on usage
Antenna Internal
Wireless range 121 m (400’)
Radio frequency 2.4 GHz ISM band
Access point USB to computer
Programs and reads unlimited number of loggers
Calibrations Factory: Traceable NIST / COFRAC - ITS-90 coefficients reside in the internal memory
®
User: Closed loop user calibration, using XpertLog Software
Certification EMC: EN 60601-1-2 / 2007
GMPF RF Health & Safety: EN 60950-1/2006
RF: EN 300440-2 v1.3.1 (2009-03)
-4-
Visit www.lives-international.com for more details. LVSC.3.1.6 - October 2019