Page 9 - Lives International Product Catalog 2019
P. 9

GMPF Bio RF







                      Real Time Temperature Data Logger -90 to +140°C




                         +    Collects and transmits real time temperature to XpertLog Software
                         +    ITS-90 calibration coefficients are stored in the internal memory
                         +    Complies with FDA 21 CFR Part 11
                         +    Network ready: multiple users, centralized, server database, active directory authentication
                         +    Calibration can be performed on site
                         +    Can be remotely activated



                   12.00mm
                     0.47”

                                                                   33.50mm/1.32″
                        89.50mm / 3.52″  101.5.00mm / 4.00″        15.50mm/0.61″










           Sensor type        4 wires  PT-100
              - flexible      480 mm (19”) PTFE
              - customized    Up to 5000 mm (195”)
              - sensor size   1.5 x 8 mm (0.06 x 0.315”)
                                             2
                                    2
              - cross section   1.77 mm  (0.0027 inch )
           Operating Range
              - body temp.    -90°C to +140°C
              - sensor temp.   -200°C to 200°C
           Accuracy           ±0.1°C from -90°C to +140°C                              Order Reference:
           Resolution         0.01°C
           Memory capacity    32,000 data points                                        GMP-Bio_RF
           Sampling rate      1 sec to 24 hours
           Transmission       Simultaneous: maximum 20 loggers
                              Sequential: unlimited
           Internal clock Drift    4 sec / 24 h @ + 23 °C
           Mechanical                                                         Standard Calibration Points
              - material      Peek 1000 - FDA compliant
              - enclosure     IP-68                                                        • -80°C
              - weight        67 g (2.36 oz)
              - diameter      33.5 mm (1.32”)
              - height        101.5 mm (4.00”)                                             •  0°C
           Power              2 x Lithium batteries, 0.35 g/each content classified
                              UN3091 under IATA DGR section 3.9.2.6 compliant with         •  140°C
                              ICAO/IATA packing instruction 970 user replaceable
           Battery lifetime   Up to 6 months, based on usage
           Antenna            Internal
           Wireless range     121 m (400’)
           Radio frequency    2.4 GHz ISM band
           Access point       USB to computer
                              Programs and reads unlimited number of loggers
           Calibrations       Factory:               Traceable NIST / COFRAC - ITS-90 coefficients reside in the internal memory
                                                                          ®
                              User:                    Closed loop user calibration, using XpertLog  Software
                                                                                                                      GMP Bio RF
           Certification      EMC:                    EN 60601-1-2 / 2007
                              RF:                       EN 300440-2 v1.3.1 (2009-03)
                              Health & Safety:      EN 60950-1/2006








                                                                                                                    -9-


                   LVSC.3.1.6 - October 2019                                        Visit www.lives-international.com for more details.