CO2 Sensor

MH-100 MicroSENS IR Incubator CO2 Sensor

Quote Request

+ Clean in place: heat-sterilization up to 190°C
+ Simple field calibration at required protocols from 0-20%
+ Temperature, pressure and humidity correction built-in
+ Self-test status continuously reported during operation
+ Metal head designed to Pharma + Lab Standards
+ Infrared dual beam CO2 sensor for accuracy and long-life

Characteristics CO2 Sensor  
Measuring gas CO2
Measurement range 0 – 20 Vol.-%
Gas supply Diffusion
Warm up time less than 1 minute (start-up)
less than 15 minutes (full spec)
Accuracy1 ± 0,2 Vol.-% ± 2 % of reading
Response time (t90) ≤ 30 s
Digital resolution 0,001 Vol.-%
Temperature dependence2 ≤ ± 0,1 Vol.-%
Pressure dependence3 ≤ ± 0,05 Vol.-%
Long term stability4 ≤ ± 0,2 Vol.-% at 5 Vol.-% / year
Humidity correction 0 … 200 hPa H2O
Supply voltage 12 – 24 VDC
Power consumption up to 2 W
Digital output RS232, Micro-Hybrid industrial protocol
Analogue output 4 – 20 mA
Operating temperature 0° C … 60° C
Humidity up to 100 % relative humidity (rH), not condensing
Storage temperature -25° C … 85° C
Maximum temperature for heat sterilization (only sensor)5 190° C

Documents

Datasheet