Xpertlog® GDP RH covers all applications between -40°C to +85°C
High efficiency: user friendly, multiple users may simultaneously share the same database
Data loggers calibration on users site
High accuracy: offers the best accuracy in its class, ±2% RH | ±0.1°C
Network ready: multiple users, centralized, server database, active directory authentication
Data integrity: FDA 21 CFR Part 11
Characteristics | GDP RH | LVSC.3.1.7 - March 2020 |
---|---|---|
Sensor type | RH: capacitive Temp: 4 wires PT-100 | |
- fixed | 20mm (0.8") | |
- temp. sensor size | 1.5 x 8 mm (0.06 x 0.315”) | |
- temp. cross section | 1.77 mm2 (0.0027 inch2) | |
Operating range | ||
- humidity | 15% to 95% RH | |
- body temp. | -40 °C to +85 °C | |
- sensor temp. | -40 °C to +85 °C | |
Accuracy | ±2.0% RH @ +23°C (15-60% RH) ±3.0%RH @ +23°C (60-95% RH) ±0.1 °C from -40 °C to +85 °C | |
Memory capacity | 16,000 data points RH / 16,000 data points temperature | |
Sampling rate | 1 sec to 24 hours | |
Internal clock drift | 4 sec / 24 h @ 23 ° | |
Mechanical | ||
- material | POM-C; FDA compliant | |
- enclosure | IP-51 | |
- weight | 45g (1.58 oz) | |
- diameter | 40 mm (1.57") | |
- height | 60 mm (2.36") | |
Power | 1 Lithium battery 0.35 g Lithium content classified UN3091 under IATA DGR section 3.9.2.6 compliant with ICAO/IATA packing instructions 970 user replaceable | |
Battery lifetime | Up to 12 months, based on usage | |
Docking station | USB connection 4, 8, 12 and 20 ports available | |
Calibrations | Factory: User: | traceable NIST/COFRAC - ITS-90 coefficients stored in the internal memory close loop calibration, using XpertLog® software |
Standard Calibration Points | 25°C / 20%RH 25°C / 45%RH 25°C / 65%RH 25°C / 75%RH Upon special request: 25°C / 90%RH | |
Certification | EMC: Health Safety: | EN 60601-1-2 / 2007 EN 60950-1 / 2006 |