Xpertlog® GMP simplifies and makes the thermal validation process more efficient:
High efficiency: easy to program and install, pre-calibration not needed
High reliability: no data loss during validation process
Network ready: multiple users, centralized, server database, active directory authentication
High accuracy: ±0.1°C from -40°C to +140°C.
Data integrity: FDA 21 CFR Part 11
Characteristics | GMP SW | LVSC.3.1.7 - March 2020 |
---|---|---|
Sensor type | 4 wires PT-100 | |
- rigid | 20mm (0.8") | |
- sensor size | 1.5 x 8 mm (0.06 x 0.315”) | |
- cross section | 1.77 mm2 (0.0027 inch2) | |
Operating range | ||
- body temp. | -40 °C to +140 °C | |
- sensor temp. | -40 °C to +140 °C | |
Accuracy | ±0.1°C from -40 to 140°C | |
Resolution | 0.01°C | |
Memory capacity | 32,000 data points temperature | |
Sampling rate | 1 sec to 24 hours | |
Internal clock drift | 4 sec / 24 h @ 23 ° | |
Mechanical | ||
- material | Peek 1000 - FDA compliant | |
- enclosure | IP-68 | |
- weight | 91g (3.21 oz) | |
- diameter | 50 mm (1.97") | |
- height | 40 mm (1.57") | |
Power | 1 Lithium battery 0.35 g Lithium content classified UN3091 under IATA DGR section 3.9.2.6 compliant with ICAO/IATA packing instructions 970 user replaceable | |
Battery lifetime | Up to 12 months, based on usage | |
Docking station | USB connection 4, 8, 12 and 20 ports available programs and reads unlimited number of loggers | |
Calibrations | Factory: User: | traceable NIST/COFRAC - ITS-90 coefficients stored in the internal memory close loop calibration, using XpertLog® software |
Standard Calibration Points | -40°C 0°C 140°C | |
Certification | EMC: Health Safety: | EN 60601-1-2 / 2007 EN 60950-1 / 2006 |