GMPF RH

Humidity Logger with Flexible T Sensor

0 to 100% RH

Standard data loggers with flexible temperature sensor and rigid humidity sensor. Easy to program and install, no data loss during validation process, multiple users can share the same database simultaneously. Remote factory calibration on users site.

Quote Request

Xpertlog® GMPF RH elevates thermal validation process to a whole new level:
Highly efficient: increase production efficiency by cutting down validation time
Highly reliable: real time data during validation process
Network ready: multiple users, centralized, server database, active directory authentication
Impressive accuracy: ±0.1°C from -90°C to +140°C.

Characteristics GMPF RH LVSC.3.1.7 - March 2020
Sensor element Humidity: capacitive
Temperature: 4 wires PT-100
- flexible 480mm (19") PTFE/SS
- customized up to 5000mm (195")
- sensor size 1.5 x 8 mm (0.06 x 0.315”)
- cross section 1.77 mm2 (0.0027 inch2)
Operating range
- humidity 0 to 100% RH non condensing
- body temp. -40 °C to +100 °C
- sensor temp. -200 °C to +200 °C
Accuracy ±2.0% RH, 0% to 100% non condensing at 20 °C to 45 °C **
±0.1 °C from -40 °C to 100 °C
Resolution 0.01% RH; 0.01°C
Memory capacity 16,000 data points humidity / 16,000 data points temperature
Sampling rate 1 sec to 24 hours
Internal clock drift 4 sec / 24 h @ 23 °
Mechanical
- material Peek 1000 - FDA compliant
- enclosure IP-51
- weight 43g (1.52 oz)
- diameter 40 mm (1.57")
- height 60 mm (2.36")
Power 1 Lithium battery
0.35 g Lithium content
classified UN3091 under IATA DGR section 3.9.2.6
compliant with ICAO/IATA packing instructions 970
user replaceable
Battery lifetime Up to 12 months, based on usage
Docking station USB connection
4, 8, 12 and 20 ports available
programs and reads unlimited number of loggers
Calibrations Factory:
User:
traceable NIST/COFRAC - ITS-90 coefficients stored in the internal memory
close loop calibration, using XpertLog® software
Standard Calibration Points 25°C / 20%RH
25°C / 45%RH
30°C / 65%RH
40°C / 75%RH
Upon special request: 45°C / 90%RH
Certification EMC:
Health Safety:
EN 60601-1-2 / 2007
EN 60950-1 / 2006

Documents

Datasheet
More in this category: « GMP RH GMP P »