Specifications for Temperature and Humidity Data Logger with Rigid Sensor Standard Data Logger GMP TH
+ Collects and stores humidity and temperature information in the internal memory+ Recover the data through the docking station using XpertLog Software
+ ITS-90 calibration coefficients are stored in the internal memory
+ Complies with FDA 21 CFR Part 11
+ Network ready: multiple users, centralized, server database, active directory authentication
Characteristics | Order Reference: GMP-TH GMP-TH-F |
LVSC.3.2.0 - January 2022 |
---|---|---|
Sensor type | Hum: capacitive / Temp: 4 wires PT-100 | |
- rigid | 20 mm (0.8”) | |
- sensor size | 1.5 x 8 mm (0.06 x 0.315”) | |
- cross section | 1.77 mm² (0.0027 inch²) | |
- body | 0% to 100% RH -40°C to +100°C |
|
- sensor | -40°C to +100°C | |
Accuracy | ±2.0% RH, 0% to 75% non condensing at 20°C to 45°C ±3.0% RH, 75% to 100% non condensing at 20°C to 45°C ±0.1°C from -40°C to +100°C |
|
Resolution | 0.01% RH; 0.01°C | |
Memory capacity | 16,000 data points humidity / 16,000 data points temperature | |
Memory type | non volatile | |
Sampling rate | 1 sec to 24 hours | |
Internal clock drift | 4 sec / 24 h @ 23°C | |
Mechanical | ||
- material | Peek 1000 - FDA compliant | |
- enclosure | IP-51 | |
- weight | 43 g (1.52 oz) | |
- diameter | 40.0 mm (1.57”) | |
- height | 60.0 mm (2.36”) | |
Power | 1 x Lithium batteries, 0.35 g/each Lithium content classified UN3091 under IATA DGR section 3.9.2.6 compliant with ICAO/IATA packing instructions 970, user replaceable |
|
Battery lifetime | Up to 12 months, based on usage | |
Docking station | USB to computer 4, 8, 12 and 20 ports available Programs and reads unlimited number of loggers |
|
Calibrations | Factory Calibrations: User Calibrations: |
traceable NIST/COFRAC - ITS-90 coefficients stored in the internal memory closed loop calibration using XpertLog® software |
Standard Calibration Points | • 25°C / 20% RH • 25°C / 45% RH • 30°C / 65% RH • 40°C / 75% RH Upon special request: • 45°C / 90% RH |
|
Certifications | EMC: Health Safety: | EN 60601-1-2 / 2007 EN 60950-1 / 2006 |