Specifications for Low Temperature Real Time Wireless Data Logger with Flexible Sensor
Usable at -120°C Wireless Data Logger GMP RF F 20R8V
+ Real time data to XpertLog Software above -60°C+ ITS-90 calibration coefficients are stored in the internal memory
+ Complies with FDA 21 CFR Part 11
+ Network ready: multiple users, centralized, server database, active directory authentication
+ Calibration can be performed on site
+ Can be remotely activated
Characteristics | Order Reference: GMP-RF-F-20R8V | LVSC.3.2.0 - January 2022 |
---|---|---|
Sensor type | 4 wires thin, teflon | |
- flexible | 150 mm (6") or custom length | |
- customized | up to 5000 mm (195”) | |
- sensor size | 1.2 x 1.6 mm (0.05 x 0.06") | |
- cross section | 0.48 mm² (0.0007 inch²) | |
- body | -90°C to +85°C -120°C up to 30 minutes |
|
- sensor | -200°C to +200°C Before exposing the logger to temperatures below -90° C, the logger must stay for at least 3 hours at ambient temperature to allow warming to ambient |
|
Accuracy | ±0.1°C from -90°C to +85°C ±0.5°C at -196°C |
|
Resolution | 0.01°C | |
Memory capacity | 32,000 data points | |
Memory type | non volatile | |
Sampling rate | 1 sec to 24 hours | |
Transmission | simultaneus: maximum 20 loggers sequential: unlimited |
|
Internal clock drift | 4 sec / 24 h @ 23°C | |
Mechanical | ||
- material | Peek 1000 - FDA compliant Screws: 316L |
|
- enclosure | IP-64 | |
- weight | 230 g (8.11 oz) | |
- length | 126 mm (4.96”) | |
- width - height |
52.25 mm (2.06”) 53.5 mm (2.09”) |
|
Power | Lithium battery user replaceable |
|
Battery lifetime | Up to 6 months, based on usage | |
Antenna | Internal | |
Wireless range | up to 121 m (400 feet) | |
Radio frequency | 2.4 Ghz ISM band | |
Access point | USB to computer Programs and reads unlimited number of loggers |
|
Calibrations | Factory Calibrations: User Calibrations: |
traceable NIST/COFRAC - ITS-90 coefficients stored in the internal memory closed loop calibration using XpertLog® software |
Standard Calibration Points | • -196°C • -80°C • 0°C • 85°C |
|
Certifications | EMC: Health Safety: RF: FCC compliant CE certified | EN 60601-1-2 / 2007 EN 60950-1 / 2006 EN 300440-2 v1.3.1 (2009-03) |