Xpertlog® GMPF Mini RF elevates thermal validation process to a whole new level:
Highly efficient: increase production efficiency by cutting down validation time
Highly reliable: real time data during validation process
Network ready: multiple users, centralized, server database, active directory authentication
Impressive accuracy: ±0.2°C from -60°C to +85°C.
Characteristics | GMPF Mini RF | LVSC.3.1.7 - March 2020 |
---|---|---|
Sensor element | 4 wires PT-100 | |
- rigid | 0.8”(20 mm) | |
- sensor size | 1.2 x 2.8 mm (0.05 x 0.11”) | |
- cross section | 1.13 mm2 (0.00175 inch2) | |
Operating Range | ||
- body temp. | -60 °C to +85 °C | |
- senzor temp. | -80 °C to +140 °C | |
Accuracy | ±0.2°C from -60 to 85°C | |
Memory capacity | 32,000 data points | |
Sampling rate | 1 sec to 24 hours | |
Internal clock drift | 4 sec / 24 h @ 23°C | |
Logger update rate | 5 sec to 24 h - user configurable | |
Mechanical | ||
- material | Peek 1000 - FDA compliant | |
- enclosure | IP-68 | |
- weight | - | |
- diameter | 36 mm | |
- height | 43 mm | |
- comments | fits in vials with a minimum 22 mm neck diameter | |
Power | Lithium batteries, user replaceable | |
Battery lifetime | Up to 1 month, based on usage | |
Wireless range | 40 feet (12 m) line of sight | |
Radio frequency | 2.4 Ghz ISM band | |
Calibrations | Factory: User: | traceable NIST/COFRAC - ITS-90 coefficients stored in the internal memory close loop calibration, using XpertLog® software |
Certification | FMC: RF: Health Safety: | EN 60601-1-2 / 2007 EN 300440-2 v1.3.1 (2009-03) EN 60950-1 / 2006 |