Features | |
---|---|
Purpose | Interface XpertLog® real time data loggers with the SCADA/HMI of an equipment Display XpertLog® real time data on SCADA / HMI |
Applications | Interface SCADA/HMI of autoclaves, freeze dryers, etc. with XpertLog® real time data loggers inside the equipment. |
Scope | Help users to meet FDA requirements with regards to Continued Process Verification Gain ongoing assurance during routine production that the process remains in a state of control Minimize time needed for parametric release of a sterile load after sterilization cycle Display real time behavior of the equipment using a third party system - XpertLog® Detect real time anomalies of product probes and allow users to take action |
Functionality | Collect temperatures/pressures from XpertLog® real time data loggers and display simultaneously with the product probes on the SCADA/HMI |
Technical Specifications | |
---|---|
Minimum recommended acquisition interval: | 10 sec |
Maximum number of loggers: | 20 |
Accuracy: | Temperature: ± 0.1°C Pressure: ± 10 mbar @ 2 bar abs. |
Data integrity: | 21 CFR Part 11 compliant |
PLC compatibility: | Any PLC model |
Measurement range: | Temperature: -60 to 140°C Pressure: 0.01 to 5 bar abs. |
Memory capacity: | 32,000 data points -16,000 p / 16,000 temp. |
Battery lifetime: | Up to 6 months - based on usage |
Internal clock drift: | 4 sec / 24 h @ +23°C |
Access Point: | USB connection |
Material: | Peek 1 - FDA compliant |
Radio frequency: | 2.4 GHz ISM band |
Enclosure: | IP-68 |
Calibrations: | Factory: Traceable NIST / COFRAC - ITS-90 coefficients stored in the internal memory User: Close loop calibration, using XpertLog® software |
Certification: | EMC: EN 60601-1-2 / 2007 RF: EN 300440-2 v1.3.1 (2009-03) Health Safety: EN 60950-1/2006 |