GMPF RF 2R 4V

4x2R Vial Size Temperature Data Logger (wireless, real time)

-50 to +85 °C

Xpertlog® GMPF RF 2R 4V small size temperature data logger designed to fit between 2R vials.

Quote Request

Xpertlog® GMPF RF 2R 4V small size temperature data logger designed to fit between 2R vials
Highly efficient: increase production efficiency by cutting down validation time
Highly reliable: real time data during validation process
Impressive accuracy: ±0.1°C from -50°C to +85°C.
Network ready: multiple users, centralized, server database, active directory authentication
Loggers calibration: can be performed on users site by Lives International
Remote activation, SCADA connectable

Characteristics GMPF RF 2R 4V LVSC.3.1.7 - March 2020
Sensor element 4 wires thin, teflon
- flexible 150 mm (6”) or custom length
- sensor size 1.2 x 2.8 mm (0.05 x 0.11”)
- cross section 1.13 mm2 (0.00175 inch2)
Operating range
- body temp. -50 °C to +85 °C
- sensor temp. -90 °C to +140 °C
Accuracy ±0.1°C from -50 to +85°C
Resolution 0.01°C
Memory capacity 32,000 data points
Sampling rate 1 sec to 24 hours
Transmission
- simultaneus: maximum 20 loggers
- sequential: unlimited
Internal clock drift 4 sec / 24 h @ 23°C
Mechanical
- material Peek 1000 - FDA compliant
Screws: 316L
- enclosure IP-68
- weight 54g (1.91 oz)
- diameter 72 mm (2.84")
- height 18 mm (0.70")
Power 2 x Lithium batteries
user replaceable
Battery lifetime Up to 6 months, based on usage
Antenna internal
Wireless range 121 m (400 feet)
Radio frequency 2.4 Ghz ISM band
Access point USB to computer
Programs and reads unlimited number of loggers
Calibrations Factory:
User:
traceable NIST/COFRAC - ITS-90 coefficients stored in the internal memory
close loop calibration, using XpertLog® software
Standard Calibration Points -50°C
0°C
85°C
Certification EMC:
RF:
Health Safety:
EN 60601-1-2 / 2007
EN 300440-2 v1.3.1 (2009-03)
EN 60950-1 / 2006

Documents

Datasheet