GMP RH RF

Humidity Data Logger (wireless, real time)

0 to 100% RH non condensing

Real time data logger for monitoring temperature. Highly efficient (quick validation time), highly reliable (real time data), impressive accuracy (±0.1°C)

Quote Request

Xpertlog® GMP RH RF temperature and humidity data logger with real time data transmission
Highly efficient: increase production efficiency by cutting down validation time
Highly reliable: real time data during validation process
Network ready:strong> multiple users, centralized, server database, active directory authentication
Impressive accuracy: ±2.0% RH @ 20°C to 45°C | ±0.1°C from -90°C to +140°C.

Characteristics GMP RH RF LVSC.3.1.7 - March 2020
Sensor type Humidity: capacitive
Temperature: 4 wires PT-100
- rigid 20 mm (0.8")
- temp. sensor size 1.5 x 8 mm (0.06 x 0.315”)
- temp. sensor cross section 1.77 mm2 (0.0027 inch2)
Operating range
- humidity 0 to 100% RH non condensing
- body temp. -40 °C to 100 °C
- sensor temp. -40 °C to 100 °C
Accuracy ±2.0% RH, 0% to 100% non condensing at 20 °C to 45 °C (custom points available at request)
±0.1 °C from -40 °C to 100 °C
Resolution 0.01% RH; 0.01 °C
Memory capacity 16,000 data points humidity / 16,000 data points temperature
Sampling rate 1 sec to 24 hours
Transmission
- simultaneus: maximum 20 loggers
- sequential: unlimited
Internal clock drift 4 sec / 24 h @ 23°C
Mechanical
- material Peek 1000 - FDA compliant
- enclosure IP-51
- weight 95g (3.35 oz)
- diameter 65 mm (2.56”)
- height 62 mm (2.44”)
Power 2 x Lithium batteries
0.35 g/each Lithium content
classified UN3091 under IATA DGR section 3.9.2.6
compliant with ICAO/IATA packing instructions 970
user replaceable
Battery lifetime Up to 6 months, based on usage
Antenna Internal
Wireless range 121 m (400 feet)
Radio frequency 2.4 Ghz ISM band
Access point USB to computer
Programs and reads unlimited number of loggers
Calibrations Factory:
User:
traceable NIST/COFRAC - ITS-90 coefficients stored in the internal memory
close loop calibration, using XpertLog® software
Standard Calibration Points 25°C / 20%RH
25°C / 45%RH
30°C / 65%RH
40°C / 75%RH
Upon special request: 45°C / 90%RH
Certification EMC:
RF:
Health Safety:
EN 60601-1-2 / 2007
EN 300440-2 v1.3.1 (2009-03)
EN 60950-1 / 2006

Documents

Datasheet