Xpertlog® GMP RF elevates thermal validation process to a whole new level
Highly efficient: increase production efficiency by cutting down validation time
Highly reliable: real time data during validation process
Network ready: multiple users, centralized, server database, active directory authentication
Impressive accuracy: ±0.1°C from -90°C to +140°C.
Characteristics | GMP RF | LVSC.3.1.7 - March 2020 |
---|---|---|
Sensor element | 4 wires PT-100 | |
- rigid | 20 mm (0.8”) / 150 mm (6”) | |
- sensor size | 1.5 x 8 mm (0.06 x 0.315”) | |
- cross section | 1.77 mm2 (0.0027 inch2) | |
Operating Range | ||
- standard | -90 °C to +140 °C | |
- extended | -90 °C to +350 °C (with thermal shield) | |
Accuracy | ±0.1°C from -90 to 140°C | |
Memory capacity | 32,000 data points | |
Sampling rate | 1 sec to 24 hours | |
Internal clock drift | 4 sec / 24 h @ 23°C | |
Logger update rate | 5 sec to 24 h - user configurable | |
Mechanical | ||
- material | Peek 1000 - FDA compliant | |
- enclosure | IP-68 | |
- weight | 115g (4.06 oz) | |
- diameter | 65 mm (2.56”) | |
- height | 55 mm (2.17”) | |
Power | 2 x Lithium batteries 0.35 g/each Lithium content classified UN3091 under IATA DGR section 3.9.2.6 compliant with ICAO/IATA packing instructions 970 user replaceable |
|
Battery lifetime | Up to 6 months, based on usage | |
Wireless range | 121 m (400') line of sight | |
Radio frequency | 2.4 Ghz ISM band | |
Calibrations | Factory: User: | traceable NIST/COFRAC - ITS-90 coefficients stored in the internal memory close loop calibration, using XpertLog® software |
Certification | FMC: RF: Health Safety: | EN 60601-1-2 / 2007 EN 300440-2 v1.3.1 (2009-03) EN 60950-1 / 2006 |