XpertVal 60

XpertVal 60

-200 to +1200°C

60 Channels (3x20) thermocouple based system for thermal mapping and validation.

Quote Request

XpertVal® is a conventional thermocouple based system for thermal mapping and validation
Highly efficient: Tc wires are installed on 3x20 channels module with cold junction compensation
Highly reliable: rugged data logger used in industrial harsh environment, real time data display
Accurate: high accuracy of Tc after pre-calibration (±0.2°C)
Data integrity: FDA 21 CFR Part 11 software

Characteristics XpertVal® LVSC.3.1.7 - March 2020
Sensors Thermocouples
Sensor length -standard: 10 m (33 feet)
Sensor type all type of thermocouples, most used: T, J, K
Temperature range - teflon: -200°C to +200°C
- kapton: -200°C to +400°C
Accuracy ±0.2°C after pre-calibration
Memory capacity no limitation, data is stored in computer
backup memory: USB storage device compatible
Tc Tip teflon: teflon tip to prevent moisture on the wire
kapton: no tip
Sampling rate typical: 10 sec. for 20 channels (high resolution)
fast scan: 2 sec. for 20 channels (low resolution)
Software 21 CFR Part 11
Module the module accepts several inputs: 4 - 20 mA / 0-10 V / RTD (PT-100 2, 3 or 4 wires)
Close loop calibration always needed when using thermocouples
system use temperature block and reference probe, software controlled
reports generated automatically by the software
Reference probe IRTD
Hart 1502
Dry blocks PTC 125: -90°C to +125°C
RTC 157: -45°C to +155°C
RTC 159: -100°C to +155°C
RTC 700 : +23°C to +700°C
Fluke 9190: -90°C to +140°C
Fluke 9170: -45°C to +140°C
Fluke 9172: +35°C to +425°C
Calibrations Factory: NIST/COFRAC
User calibration

Documents

Datasheets