Xpertlog® GMPF RF 30H temperature data logger with 30H vial shape:
Highly efficient: increase production efficiency by cutting down validation time
Highly reliable: real time data during validation process
Impressive accuracy: ±0.1°C from -90°C to +140°C.
Network ready: multiple users, centralized, server database, active directory authentication
Loggers calibration: can be performed on users site by Lives International
Remote activation, SCADA connectable
Characteristics | GMPF RF 30H | LVSC.3.1.7 - March 2020 |
---|---|---|
Sensor type | 4 wires PT-100 2 wires PT-100 | |
- flexible | 4 wires: 480 mm (19") 2 wires: 150 mm (6") or custom length | |
- sensor size | 4 wires: 1.5 x 8 mm (0.06 x 0.315”) 2 wires: 1.2 x 2.8 mm (0.05 x 0.11”) | |
- cross section | 4 wires: 1.77 mm2 (0.0027 inch2) 2 wires: 1.13 mm2 (0.00175 inch2) | |
Operating range | ||
- body temp. | -90 °C to +140 °C | |
- 4 wires temp. | -200 °C to +200 °C | |
- 2 wires temp. | -90 °C to +140 °C | |
Accuracy | ||
- 4 wires | ±0.1°C from -90 to +140°C | |
- 2 wires | ±0.2°C from -90 to +140°C | |
Resolution | 0.01°C | |
Memory capacity | 32,000 data points | |
Sampling rate | 1 sec to 24 hours | |
Transmission | ||
-simultaneous | maximum 20 loggers | |
-sequential | unlimited | |
Internal clock drift | 4 sec / 24 h @ 23°C | |
Mechanical | ||
- material | Peek 1000 - FDA compliant | |
- enclosure | IP-68 | |
- weight | 50g (1.76 oz) | |
- diameter | 37 mm (1.46") | |
- height | 62.81 mm (2.47") | |
Power | 2 Lithium batteries 0.35 g/each Lithium content classified UN3091 under IATA DGR section 3.9.2.6 compliant with ICAO/IATA packing instructions 970 user replaceable | |
Battery lifetime | Up to 6 months, based on usage | |
Antenna | internal | |
Wireless range | 121 m (400 feet) | |
Radio frequency | 2.4 Ghz ISM band | |
Access point | USB to computer Programs and reads unlimited number of loggers | |
Calibrations | Factory: User: | traceable NIST/COFRAC - ITS-90 coefficients stored in the internal memory close loop calibration, using XpertLog® software |
Standard Calibration Points | -80°C 0°C 140°C | |
Certification | EMC: RF: Health Safety: | EN 60601-1-2 / 2007 EN 300440-2 v1.3.1 (2009-03) EN 60950-1 / 2006 |