GMPF P RF

Pressure Data Logger (flexible T sensor)

0.01 to 5 BarA

Real time data logger for monitoring pressure, with flexible temperature sensor. Highly efficient (quick validation time), highly reliable (real time data), impressive accuracy (±0.1°C)

Quote Request

Xpertlog® GMP P RF elevates thermal validation process to a whole new level:
Highly efficient: increase production efficiency by cutting down validation time
Highly reliable: real time data during validation process
Impressive accuracy: ±10 mbar ≤ 3 BarA @ 130°C / ±0.1°C from -40 °C to 140 °C
Network ready: multiple users, centralized, server database, active directory authentication
Loggers calibration: can be performed on users site by Lives International
Remote activation, SCADA connectable

Characteristics GMP P RF LVSC.3.1.7 - March 2020
Sensor type Absolute pressure
Temperature: 4 wires PT-100
- rigid 20 mm (0.8")
- sensor size 1.5 x 8 mm (0.06 x 0.315”)
- cross section 1.77 mm2 (0.0027 inch2)
Operating range
- pressure 0.01 to 5 BarA
- body. temp -40 °C to 140 °C
- sensor temp. -40 °C to 140 °C
Accuracy ±10 mbar ≤ 3 BarA @ 130°C
±15 mbar > 3 BarA @ 130°C
±15 mbar ≤ 3 BarA @ 23°C
±30 mbar > 3 BarA @ 23°C
±0.1°C from -40 to 140°C
Resolution 0.1 mbar; 0.01 °C
Memory capacity 16,000 data points pressure / 16,000 data points temperature
Sampling rate 1 sec to 24 hours
Transmission
- simultaneous: maximum 20 loggers
- sequential: unlimited
Internal clock drift 4 sec / 24 h @ 23°C
Mechanical
- material Peek 1000 - FDA compliant
- enclosure IP-68
- weight 122g (4.30 oz)
- diameter 64.5 mm (2.54")
- height 65.2 mm (2.57")
Power 2 x Lithium batteries
0.35 g/each Lithium content
classified UN3091 under IATA DGR section 3.9.2.6
compliant with ICAO/IATA packing instructions 970
user replaceable
Battery lifetime Up to 6 months, based on usage
Antenna Internal
Wireless range 121 m (400 feet)
Radio frequency 2.4 Ghz ISM band
Access point USB to computer
Programs and reads unlimited number of loggers
Calibrations Factory:
User:
traceable NIST/COFRAC-ITS-90 coefficients stored in the internal memory
close loop calibration, using XpertLog® software
Standard Calibration Points hysteresis at
23°C and 130°C:
0.05 BarA
0.50 BarA
1.00 BarA
1.50 BarA
2.00 BarA
2.50 BarA
3.00 BarA
3.50 BarA
4.00 BarA
4.50 BarA
5.00 BarA
Certification EMC:
RF:
Health Safety:
EN 60601-1-2 / 2007
EN 300440-2 v1.3.1 (2009-03)
EN 60950-1 / 2006

Documents

Datasheet