Xpertlog® GMP5F RF HT elevates thermal validation process to a whole new level:
Highly efficient: increase production efficiency by cutting down validation time
Highly reliable: real time data during validation process
Impressive accuracy: ±0.1°C from -40°C to +140°C; ±0.4°C from 141°C to +350°C
Network ready: multiple users, centralized, server database, active directory authentication
Remote activation, SCADA connectable
Characteristics | GMP5F RF HT | LVSC.3.1.7 - March 2020 |
---|---|---|
Sensor type | PT-100 | |
- flexible | customizable length starting at 2 m (78.74”) | |
- sensor size | 1.5 x 8 mm (0.06 x 0.315”) | |
- cross section | 1.77 mm2 (0.0027 inch2) | |
Operating range | ||
- temperature range | -40 °C to +350 °C | |
- duration at high temperature | 5 minutes at +350°C 15 minutes at +325°C 30 minutes at +300°C | |
Accuracy | ±0.1°C from -40 to 140°C ±0.4°C from 141 to 350°C | |
Resolution | 0.01°C | |
Memory capacity | 32,000 data points (6400 data points/sensor) | |
Sampling rate | 1 sec to 24 hours | |
Transmission | simultaneous sequential | |
Internal clock drift | 4 sec / 24 h @ 23°C | |
Mechanical | ||
- material | Outer casing: Stainless Steel 316L | |
- enclosure | IP-51 | |
- weight | 4.5 kg (158.7 oz) | |
- size | 368x215x50 mm (14.49x8.46x1.97”) | |
Power | 1 Lithium battery | |
Battery lifetime | Up to 4 months, based on usage | |
Antenna | Internal | |
Wireless range | 121 m (400 feet) | |
Radio frequency | 2.4 Ghz ISM band | |
Access point | USB to computer Programs and reads unlimited number of loggers | |
Calibrations | Factory: User: | traceable NIST/COFRAC - ITS-90 coefficients stored in the internal memory close loop calibration, using XpertLog® software |
Standard Calibration Points | -40°C 0°C 140°C 350°C | |
Certification | EMC: RF: Health Safety: | EN 60601-1-2 / 2007 EN 300440-2 v1.3.1 (2009-03) EN 60950-1 / 2006 |