GMPF CO2 RF

CO2 Data Logger (flexible T sensor)

≤ 20% CO2 / -200 to +200°C

GMPF CO2 RF collects real time CO2, temperature and pressure values and sends the data to XpertLog® Software. With an accuracy of ±0.5% CO2 and ±0.1°C, GMPF CO2 RF allows you to measure CO2 up to 20% and temperature from -200 to +200°C.

Quote Request

Xpertlog® GMPF CO2 RF collects and transmits real time CO2 and temperature to XpertLog Software
Highly efficient: increase production efficiency by cutting down validation time
Highly reliable: real time data during validation process
Impressive accuracy: ±0.1°C from -90°C to +140°C; ±0.5% CO2.
Network ready: multiple users, centralized, server database, active directory authentication
Loggers calibration: can be performed on users site by Lives International
Remote activation, SCADA connectable

Characteristics GMPF CO2 RF LVSC.3.1.7 - March 2020
Sensor type CO2 - IR / Temp: 4 wires PT-100
- flexible 480 mm (19”)
- customized up to 5000 mm (195”)
- temp. sensor size 1.5 x 8 mm (0.06 x 0.315”)
- temp. cross section 1.77 mm2 (0.0027 inch2)
Operating range
- CO2 0 to 20% CO2
- body temp. 0 °C to +50 °C
- sensor temp. -200 °C to +200 °C
- pressure 300 to 1100 mBar (built-in pressure compensation)
Accuracy
- temperature ±0.1°C from -90 to 140°C
- CO2 ±0.5% CO2
Resolution 0.01% CO2; 0.01°C
Memory capacity 16,000 data points CO2 / 16,000 data points temperature
Sampling rate 2 min to 24 hours
Transmission simultaneous: maximum 20 loggers
sequential: unlimited
Internal clock drift 4 sec / 24 h @ 23°C
Mechanical
- material Peek 1000 - FDA compliant
- enclosure IP-30
- weight 153g (5.39 oz)
- diameter 64.5 mm (2.53")
- height 52 mm (2.04")
Power 2 x Lithium batteries
0.35 g/each Lithium content
classified UN3091 under IATA DGR section 3.9.2.6
compliant with ICAO/IATA packing instructions 970
user replaceable
Battery lifetime Up to 6 months, based on usage
Antenna Internal
Wireless range 121 m (400 feet)
Radio frequency 2.4 Ghz ISM band
Access point USB to computer
Programs and reads unlimited number of loggers
CalibrationsFactory:
User:
traceable NIST/COFRAC - ITS-90 coefficients stored in the internal memory
close loop calibration, using XpertLog® software
Certification EMC:
RF:
Health Safety:
EN 60601-1-2 / 2007
EN 300440-2 v1.3.1 (2009-03)
EN 60950-1 / 2006

Documents

Datasheet