Standard Pressure Data Logger

GMPF P

0.01 to 5.0 BarA
-40°C to +140°C

Standard data logger for recording pressure and temperature values, with flexible temperature sensor. Easy to program and install, no data loss during validation process, multiple users can share the same database simultaneously. Remote factory calibration on users site.

Quote Request

Specifications for Standard Pressure Data Logger GMPF P

+ Collects and stores pressure and temperature information in the internal memory
+ Recover the data through the docking station using XpertLog Software
+ ITS-90 coefficients are stored in the internal memory
+ Complies with FDA 21 CFR Part 11
+ Network ready: multiple users, centralized, server database, active directory authentication
Characteristics Order Refference: GMPF-P LVSC.3.1.8 - May 2020
Sensor type Absolute pressure / Temp: 4 wires PT-100
- flexible 480 mm (19”)
- customized up to 5000 mm (195”)
- sensor size 1.5 x 8 mm (0.06 x 0.315”)
- cross section 1.77 mm² (0.0027 inch²)
- body 0.01 to 5.0 BarA
-40°C to +140°C
- sensor -40°C to +140°C
Accuracy ±10 mbar ≤ 3 BarA @ 130°C; ±15 mbar > 3 BarA @ 130°C; ±15 mbar ≤ 3 BarA @ 23°C; ±30 mbar > 3 BarA @ 23°C; ±0.1°C from -40 to +140°C
Resolution 0.1 mbar; 0.01°C
Memory capacity 16,000 data points pressure / 16,000 data points temperature
Memory type non volatile
Sampling rate 1 sec to 24 hours
Internal clock drift 4 sec / 24 h @ 23°C
Mechanical
- material Peek 1000 - FDA compliant
- enclosure IP-51
- weight 43 g (1.52 oz)
- diameter 40.0 mm (1.57”)
60.0 mm (2.36”)
Power 1 x Lithium batteries, 0.35 g/each Lithium content
classified UN3091 under IATA DGR section 3.9.2.6
compliant with ICAO/IATA packing instructions 970,
user replaceable
Battery lifetime Up to 12 months, based on usage
Docking station USB to computer
4, 8, 12 and 20 ports available
Programs and reads unlimited number of loggers
Calibrations Factory Calibrations:
User Calibrations:
traceable NIST/COFRAC - ITS-90 coefficients stored in the internal memory
close loop calibration using XpertLog® software
Standard Calibration Points hysteresis at 23°C and 130°C:
• 0.05 BarA
• 0.50 BarA
• 1.00 BarA
• 1.50 BarA
• 2.00 BarA
• 2.50 BarA
• 3.00 BarA
• 3.50 BarA
• 4.00 BarA
• 4.50 BarA
• 5.00 BarA
Certifications EMC:
Health Safety:
EN 60601-1-2 / 2007
EN 60950-1 / 2006

Related Models for Standard Pressure Data Logger GMPF P

Documents for Standard Pressure Data Logger GMPF P

GMPF P Datasheet