Xpertlog® GDPF NBF covers all the applications between -40°C to +85°C
High efficiency: user friendly, multiple users may simultaneously share the same database
Network ready: multiple users, centralized, server database, active directory authentication
Data loggers calibration on users site
High accuracy: ±0.1°C from -40°C to +85°C.
Data integrity: FDA 21 CFR Part 11
Characteristics | GDPF NBF | LVSC.3.1.7 - March 2020 |
---|---|---|
Sensor element | 4 wires PT-100 | |
- flexible | 480mm (19") | |
- customized | up to 5000mm (195") | |
- sensor size | 1.5 x 8 mm (0.06 x 0.315”) | |
- cross section | 1.77 mm2 (0.0027 inch2) | |
Operating range | ||
- body temp. | -40 °C to +85 °C | |
- sensor temp. | -40 °C to +85 °C | |
Accuracy | ±0.1 °C from -40 °C to +85 °C | |
Resolution | 0.01 °C | |
Memory capacity | 32,000 data points | |
Sampling rate | 1 sec to 24 hours | |
Internal clock drift | 4 sec / 24 h @ 23 ° | |
Mechanical | ||
- material | POM-C; FDA compliant | |
- enclosure | IP-67 | |
- weight | 44 g (1.55 oz) | |
- diameter | 40 mm (1.57") | |
- height | 50 mm (1.97") | |
Power | 1 Lithium battery 0.35 g Lithium content classified UN3091 under IATA DGR section 3.9.2.6 compliant with ICAO/IATA packing instructions 970 user replaceable | |
Battery lifetime | Up to 12 months, based on usage | |
Docking station | USB to computer 4, 8, 12 and 20 ports available programs and reads unlimited number of loggers | |
Calibrations | Factory: User: | traceable NIST/COFRAC - ITS-90 coefficients stored in the internal memory close loop calibration, using XpertLog® software |
Standard Calibration Points | -40°C 0°C 85°C | |
Certification | EMC: Health Safety: | EN 60601-1-2 / 2007 EN 60950-1 / 2006 |