Xpertlog® GMP elevates thermal validation process to a whole new level:
Highly efficient: increase production efficiency by cutting down validation time
Highly reliable: real time data during validation process
Network ready: multiple users, centralized, server database, active directory authentication
Impressive accuracy: ±10 mbar ≤ 3 BarA @ 130°C / ±0.1°C from -40 °C to 140 °C
Characteristics | GMPF P | LVSC.3.1.7 - March 2020 |
---|---|---|
Sensor type | absolute pressure 4 wires PT-100 | |
- flexible | - 480mm (19") PTFE/SS | |
- customized | - up to 5000mm (195") | |
- temp. sensor size | 1.5 x 8 mm (0.06 x 0.315”) | |
- temp. sensor cross section | 1.77 mm2 (0.0027 inch2) | |
Operating range | ||
- pressure | 0.01 to 5 BarA | |
- body temp. | -40 °C to +140 °C | |
- sensor temp. | -200 °C to +200 °C | |
Accuracy | ±10 mbar ≤ 3 BarA @ 130°C ±15 mbar > 3 BarA @ 130°C ±15 mbar ≤ 3 BarA @ 23°C ±30 mbar > 3 BarA @ 23°C ±0.1°C from -40 to 140°C | |
Resolution | 0.1 mbar; 0.01°C | |
Memory capacity | 16,000 data points pressure / 16,000 data points temperature | |
Sampling rate | 1 sec to 24 hours | |
Internal clock drift | 4 sec / 24 h @ 23 ° | |
Mechanical | ||
- material | Peek 1000 - FDA compliant | |
- enclosure | IP-68 | |
- weight | 77g (2.72 oz) | |
- diameter | 40 mm (1.57") | |
- height | 96 mm (3.78") | |
Power | 1 Lithium battery 0.35 g Lithium content classified UN3091 under IATA DGR section 3.9.2.6 compliant with ICAO/IATA packing instructions 970 user replaceable | |
Battery lifetime | Up to 12 months, based on usage | |
Docking station | USB connection 4, 8, 12 and 20 ports available programs and reads unlimited number of loggers | |
Calibrations | Factory: User: | traceable NIST/COFRAC - ITS-90 coefficients stored in the internal memory close loop calibration, using XpertLog® software |
Standard Calibration Points | hysteresis at 23°C and 130°C: | 0.05 BarA 0.50 BarA 1.00 BarA 1.50 BarA 2.00 BarA 2.50 BarA 3.00 BarA 3.50 BarA 4.00 BarA 4.50 BarA 5.00 BarA |
Certification | EMC: Health Safety: | EN 60601-1-2 / 2007 EN 60950-1 / 2006 |