Specifications for Real Time Wireless Temperature and Pressure Data Logger with Flexible Sensor Wireless Data Logger GMP RF TP F
+ Collects and transmits real time pressure and temperature to XpertLog Software+ ITS-90 calibration coefficients are stored in the internal memory
+ Complies with FDA 21 CFR Part 11
+ Network ready: multiple users, centralized, server database, active directory authentication
+ Can be remotely activated
Characteristics | Order Reference: GMP-RF-TP-F | LVSC.3.2.0 - January 2022 |
---|---|---|
Sensor type | Absolute pressure / Temp: 4 wires PT-100 | |
- flexible | 430 mm (17”) | |
- customized | up to 5000 mm (195”) | |
- sensor size | 1.5 x 8 mm (0.06 x 0.315”) | |
- cross section | 1.77 mm² (0.0027 inch²) | |
- body | 0.01 BarA to 5 BarA -40°C to +140°C |
|
- sensor | -200°C to +200°C | |
Accuracy | ±10 mbar ≤ 3 BarA @ 130°C; ±15 mbar > 3 BarA @ 130°C; ±15 mbar ≤ 3 BarA @ 23°C; ±30 mbar > 3 BarA @ 23°C; ±0.1°C from -40°C to +140°C | |
Resolution | 0.1mbar; 0.01°C | |
Memory capacity | 16,000 data points pressure / 16,000 data points temperature | |
Memory type | non volatile | |
Sampling rate | 1 sec to 24 hours | |
Transmission | simultaneus: maximum 20 loggers sequential: unlimited |
|
Internal clock drift | 4 sec / 24 h @ 23°C | |
Mechanical | ||
- material | Peek 1000 - FDA compliant | |
- enclosure | IP-68 | |
- weight | 124 g (4.37 oz) | |
- diameter | 64.50 mm (2.54”) | |
- height | 51.0 mm (2.01”) / 72.0 mm (2.83”) | |
Power | 2 x Lithium batteries, 0.35 g/each Lithium content classified UN3091 under IATA DGR section 3.9.2.6 compliant with ICAO/IATA packing instructions 970, user replaceable |
|
Battery lifetime | Up to 6 months, based on usage | |
Antenna | Internal | |
Wireless range | 121 m (400 feet) | |
Radio frequency | 2.4 Ghz ISM band | |
Access point | USB to computer Programs and reads unlimited number of loggers |
|
Calibrations | Factory Calibrations: User Calibrations: |
traceable NIST/COFRAC - ITS-90 coefficients stored in the internal memory closed loop calibration using XpertLog® software |
Standard Calibration Points | hysteresis at 23°C and 130°C: • 0.05 BarA • 0.50 BarA • 1.00 BarA • 1.50 BarA • 2.00 BarA • 2.50 BarA • 3.00 BarA • 3.50 BarA • 4.00 BarA • 4.50 BarA • 5.00 BarA |
|
Certifications | EMC: Health Safety: RF: | EN 60601-1-2 / 2007 EN 60950-1 / 2006 EN 300440-2 v1.3.1 (2009-03) |