Page 25 - Lives International Product Catalog 2019
P. 25

GMPF  LP






                            Standard Temperature Data Logger  -90 to +140°C










                                  +    Collects and stores temperature information in the internal memory
                                  +    Recover the data through the docking station using XpertLog Software
                                  +    ITS-90 calibration coefficients are stored in the internal memory
                                  +    Complies with FDA 21 CFR Part 11
                                  +    Network ready: multiple users, centralized, server database, active directory authentication
                                  +    Calibration can be performed on site





                             7.00 mm  0.28″

                                                               15.00mm / 0.59″
                                                                40.00mm / 1.57″″
                            28.00 mm     1.10″  35mm / 1.38″





              Sensor type        4 wires  PT-100
                 - flexible      480 mm (19”)
                 - customized    Up to 5000 mm (195”)
                 - sensor size   1.5 x 8 mm (0.06 x 0.315”)                                 Order Reference:
                 - cross section   1.77 mm  (0.0027 inch )
                                       2
                                                 2
              Operating Range                                                               GMPF-LP
                 - body temp.    -90°C to +140°C
                 - sensor temp.   -200°C to +200°C
              Accuracy           ±0.1°C from -90 to +140°C
              Resolution         0.01°C
              Memory capacity    32,000 data points
              Sampling rate      1 sec to 24 hours                                Standard Calibration Points
              Internal clock Drift    4 sec / 24 h @ + 23 °C
              Mechanical                                                                • -80°C
                 - material      Peek 1000 - FDA compliant
                 - enclosure     IP-68                                                  • 0°C
                 - weight        44 g (1.55 oz)
                 - diameter      40.0 mm (1.57”)                                        • 140°C
                 - height        35.0 mm (1.38”)
              Power              1 x Lithium battery, 0.35 g Lithium content classified
                                 UN3091 under IATA DGR section 3.9.2.6 compliant with
                                 ICAO/IATA packing instruction 970 user replaceable
              Battery lifetime   Up to 12 months, based on usage
              Docking station    USB to computer
                                 4, 8, 12 and 20 ports available
                                 Programs and reads unlimited number of loggers
              Calibrations       Factory:               Traceable NIST / COFRAC - ITS-90 coefficients reside in the internal memory
                                                                             ®
                                 User:                    Closed loop user calibration, using XpertLog  Software
              Certification      EMC:                    EN 60601-1-2 / 2007
                                                                                                                      GMPF LP
                                 Health & Safety:      EN 60950-1/2006










                                                                                                                   -25-
   20   21   22   23   24   25   26   27   28   29   30