Xpertlog® GMPF RF 10R small size temperature data logger designed to fit between 10R vials
Highly efficient: increase production efficiency by cutting down validation time
Highly reliable: real time data during validation process
Impressive accuracy: ±0.1°C from -60°C to +85°C.
Network ready: multiple users, centralized, server database, active directory authentication
Loggers calibration: can be performed on users site by Lives International
Remote activation, SCADA connectable
Characteristics | GMPF RF 10R | LVSC.3.1.7 - March 2020 |
---|---|---|
Sensor element | 4 wires thin | |
- flexible | 150 mm (6”) or custom length | |
- sensor size | 1.2 x 2.8 mm (0.05 x 0.11”) | |
- cross section | 1.13 mm2 (0.00175 inch2) | |
Operating range | ||
- body temp. | -60 °C to +85 °C | |
- sensor temp. | -80 °C to +140 °C | |
Accuracy | ±0.2°C from -60 to +85°C | |
Resolution | 0.01°C | |
Memory capacity | 32,000 data points | |
Sampling rate | 1 sec to 24 hours | |
Transmission | ||
- simultaneus: | maximum 20 loggers | |
- sequential: | unlimited | |
Internal clock drift | 4 sec / 24 h @ 23°C | |
Mechanical | ||
- material | Peek 1000 - FDA compliant | |
- enclosure | IP-68 | |
- weight | 24g (0.85 oz) | |
- diameter | 25 mm (0.98") | |
- height | 46 mm (1.81") | |
Power | 1 Lithium battery 0.35 g Lithium content classified UN3091 under IATA DGR section 3.9.2.6 compliant with ICAO/IATA packing instructions 970 user replaceable | |
Battery lifetime | Up to 2 months, based on usage | |
Antenna | internal | |
Wireless range | 121 m (400 feet) | |
Radio frequency | 2.4 Ghz ISM band | |
Access point | USB to computer Programs and reads unlimited number of loggers | |
Calibrations | Factory: User: | traceable NIST/COFRAC - ITS-90 coefficients stored in the internal memory close loop calibration, using XpertLog® software |
Standard Calibration Points | -60°C 0°C 85°C | |
Certification | EMC: RF: Health Safety: | EN 60601-1-2 / 2007 EN 300440-2 v1.3.1 (2009-03) EN 60950-1 / 2006 |