XpertVal Description
60 Channels (3x20) thermocouple based system for thermal mapping and validation.
XpertVal Main Features
+ Conventional Tc’s based system for thermal mapping and validation
+ 60 (up to 120) channels simultaneously
+ Rugged data logger used in industrial harsh environment, real time data display
+ High accuracy Tc after pre-calibration (±0.2°C)
+ Complies with FDA 21 CFR Part 11
+ Network ready: multiple users, centralized, server database, active directory authentication
Order Reference:
Lives-XpertVal60
Characteristics | Revision: LVSC.3.2.0 - January 2022 | |
---|---|---|
Measuring range | up to +1200°C | |
Sensors | Thermocouples | |
- standard: | 10 m (33 feet) | |
-customizable: | custom length | |
Sensor type | All type of thermocouples, most used : T, J, K | |
- teflon: | -200°C to +200°C | |
- kapton: | -200°C to +400°C | |
- other: | up to +1200°C | |
Accuracy | ±0.2°C after pre-calibration | |
Memory capacity | Internal memory: 20 Mb Backup memory: USB storage device compatible |
|
Resolution | 0.01°C | |
TC Tip | Teflon: teflon tip to prevent moisture on the wire Kapton: no tip |
|
Sampling rate | 120 channels - 5 sec 60 channels - 5 sec 20 channels - 2 sec 12 channels - 1 sec |
|
Software | 21 CFR Part 11 compliant | |
Module | Universal input module: 4 - 20 mA / 0-10 V / RTD (PT-100 2, 3 or 4 wires), thermocouples | |
Communication | USB XpertVal to PC LAN 2.4 GHz, 802.11b/g/n |
|
Closed loop calibration | Required when using thermocouples for thermal validation System uses temperature block and reference probe, software controlled reports generated automatically by the software |
|
Reference probes | RTD, Fluke 1502, Lives XpertRTD | |
Dry blocks | Lives: PTC 125: -90°C to +125°C / RTC 157: -45°C to +155°C / RTC 159: -100°C to +155°C RTC 700: +23°C to +700°C Fluke 9190: -90°C to +140°C; Fluke 9170: -45°C to +140°C; Fluke 9172: +35°C to +425°C |
|
Calibrations | Factory: NIST / COFRAC Users: closed loop calibration using XpertLog® software |
|
Order Reference | Lives-XpertVal60 | |
Revision: LVSC.3.2.0 - January 2022 |