MH-100 MicroSENS IR Incubator CO2 Sensor

MicroSENSE CO2 Sensor

≤ ± 0,1 Vol.-%

MH-100 MicroSENS IR Incubator CO2 Sensor

Quote Request
+ Clean in place: heat-sterilization up to 190°C
+ Simple field calibration at required protocols from 0-20%
+ Temperature, pressure and humidity correction built-in
+ Self-test status continuously reported during operation
+ Metal head designed to Pharma + Lab Standards
+ Infrared dual beam CO2 sensor for accuracy and long-life
Characteristics Order Reference: MH-100-CO2 LVSC.3.1.8 - May 2020
Measuring gas CO2 - IR
0 – 20 Vol.-% Diffusion
Warm up time < 1 minute (start-up)
< 15 minutes (full spec)
Accuracy ± 0,2 Vol.-% ± 2 % of reading
Response time (t90) ≤ 30 s
Digital resolution 0,001 Vol.-%
Temperature dependence ≤ ± 0,1 Vol.-%
Pressure dependence ≤ ± 0,05 Vol.-%
Long term stability ≤ ± 0,2 Vol.-% at 5 Vol.-% / year
Humidity correction 0 … 200 hPa H2O
Supply voltage 12 – 24 VDC
Power consumption < 2 W
Digital output RS232, Micro-Hybrid industrial protocol
Analogue output Operating temperature 0° C … 60° C
Humidity < 100 % relative humidity (rH), not condensing
Storage temperature -25° C … 85° C
Mechanical
Maximum temperature for heat sterilization (only sensor) 190° C
1 at 37° C, 1013 hPa, dry test gas, excludes calibration gas tolerance of ± 1 %
2 with compensation at 1 Vol.% … 20 Vol.% CO2 and 20° C … 60° C, 1013 hPa
3 with compensation at 600 – 1200 hPa, 37° C and 5 Vol.-% CO2
4 stability at 37° C, without heat sterilization
5 maximum humidity ≤ 1 % rH, ≥ 85° C auto standby – no CO2 measurement


MicroSENSE CO2 Sensor Documents

MicroSENSE CO2 Sensor Datasheet